MDRC logo
Explore
DE RU
← Knowledge centre

Medical software · regulatory engineering · EU market access

How to obtain CE marking for medical software

The software build and the regulatory build must reach release together

CE marking is not a logo added after development. It is the visible result of a system connecting the intended purpose, software architecture, clinical evidence, risk controls, quality processes and post-market responsibilities.

15 min read
Open the release pathway ↓

One product, two synchronized builds

A reliable application is not automatically a compliant medical device. The product and its regulatory evidence must describe the same intended purpose, the same risks, the same functionality and the same released version.

Software engineering

Build the function

Requirements, architecture, implementation, interfaces, verification, usability, cybersecurity, configuration and controlled release.

Regulatory engineering

Build the argument

Qualification, classification, risks, GSPR, clinical or performance evidence, QMS, conformity assessment and lifecycle control.

01

First decide whether the software is a medical device #

Medical Device Software is software intended to be used alone or in combination for a purpose covered by the definition of a medical device or an in vitro diagnostic medical device.

It may qualify as MDSW when it supports diagnosis or treatment decisions, monitors physiological processes, calculates patient-specific treatment, analyses medical images, interprets diagnostic data or predicts clinically relevant outcomes.

The software does not need to control a physical device or act directly on the human body. A standalone application can be a medical device in its own right when its intended purpose is medical.

  • Diagnostic or therapeutic decision support
  • Patient-specific calculations or predictions
  • Analysis of images, signals or physiological data
  • Monitoring of physiological processes
  • Interpretation of laboratory or other diagnostic data

Not every digital health product is MDSW. Software used only for scheduling, invoicing, communication, storage, archiving, lossless transfer or general wellness may fall outside the MDR and IVDR if it does not perform a medical function.

The first regulatory decision

Do not begin by selecting a risk class. Begin by deciding whether the product is a medical device, an IVD, an accessory or a non-device product.

The current European approach is explained in the European Commission’s MDCG 2019-11 rev.1 guidance. MDRC also discusses the boundary in the article on software as a medical device or an accessory.

02

Define the intended purpose before building the pathway #

The intended purpose is the foundation of the conformity assessment. It determines what the software promises, which regulation applies, what evidence is needed and what can happen if the output is wrong.

A useful intended-purpose statement identifies the medical function, users, patient population, use environment, inputs, outputs and clinical context. “Software for healthcare professionals” does not define a regulated product.

  • The medical function and clinical claim
  • The intended patient population and user
  • The use environment and workflow
  • The input data and generated output
  • The role of that output in a medical decision

The product boundary matters equally. A digital platform may combine a regulated clinical function with communication, storage or administrative modules. Those functions do not automatically share one regulatory status. The manufacturer must show which components belong to the medical device and which remain outside it.

A clear regulatory strategy prevents the marketing claims, technical architecture and conformity route from drifting apart during development.

03

Determine whether the MDR or IVDR applies #

Software generally falls under the MDR when its purpose concerns diagnosis, prevention, prediction, prognosis, monitoring, treatment or alleviation of disease without meeting the specific definition of an IVD.

Software may fall under the IVDR when it provides information by examining data derived from human specimens and that information serves an IVDR purpose, such as identifying a pathological condition, determining predisposition, predicting treatment response or monitoring therapeutic measures.

The data source is relevant but not decisive by itself. The intended purpose, the nature of the generated information and the way that information is used must be assessed together.

MDR pathway

Information or functions serving a medical purpose outside the specific scope of in vitro examination.

IVDR pathway

Information generated from the examination of specimens derived from the human body for an IVDR purpose.

A laboratory information system that only stores and transfers results may not be an IVD. Software that interprets laboratory measurements and produces patient-specific diagnostic information may be one.

The legal definitions appear in the official texts of the Medical Device Regulation and the In Vitro Diagnostic Medical Device Regulation.

04

Classify the software according to its clinical impact #

Qualification asks whether the product is a medical device. Classification asks how much regulatory control it requires.

Medical software under the MDR is commonly classified using Rule 11 of Annex VIII. Software providing information used for diagnostic or therapeutic decisions is generally Class IIa. It moves to Class IIb or Class III when an incorrect decision may lead to more serious consequences.

Software monitoring physiological processes is generally Class IIa. It may become Class IIb when it monitors vital parameters whose variation could create immediate danger. Software outside those parts of Rule 11 may remain Class I.

LOWER CONTROLClass IOther software falling outside the higher parts of Rule 11
DECISION SUPPORTClass IIaDefault class for many diagnostic, therapeutic or monitoring functions
SERIOUS IMPACTClass IIbDecisions or monitoring where failure may cause serious deterioration
CRITICAL IMPACTClass IIIDecisions whose consequences may include death or irreversible deterioration

For IVDR software, classification follows its intended purpose and the applicable IVDR rules. Independent software is classified in its own right. Software that drives or influences an IVD generally follows the class of that device.

The classification rationale should connect the intended purpose, applicable rule and foreseeable clinical consequences. Class I should never be selected merely because the product is “only software.”

05

Establish a quality system that controls the real development process #

Every manufacturer needs a quality management system appropriate to the product, risk class and activities it performs. The QMS must control more than software coding.

Product planning

Regulatory strategy, intended purpose, requirements and development controls

Evidence control

Risk management, clinical evaluation, verification, validation and usability

Operational control

Suppliers, outsourced processes, releases, changes and configuration

Lifecycle control

PMS, vigilance, cybersecurity, CAPA and continuing compliance

EN ISO 13485 is widely used to structure this system and may provide evidence of conformity within its scope. It does not replace the MDR or IVDR. The manufacturer must still demonstrate that every applicable regulatory obligation is addressed.

For software requiring Notified Body involvement, the QMS will normally be assessed as part of the conformity assessment. A polished set of procedures will not be enough if the development records show that the procedures were not followed.

06

Build lifecycle evidence, not an IEC 62304 claim #

IEC 62304 provides a recognised framework for software lifecycle processes. It addresses planning, requirements, architecture, detailed design, implementation, integration, testing, release, maintenance, configuration management and problem resolution.

It is important, but it is not the complete regulatory answer. IEC 62304 does not by itself demonstrate conformity with all General Safety and Performance Requirements. It does not replace risk management, usability engineering, clinical evidence, cybersecurity, labelling, quality management or post-market surveillance.

IEC 62304

Software lifecycle processes and controlled maintenance

ISO 14971

Medical-device risk management across the product lifecycle

IEC 62366-1

Usability engineering and control of use-related risks

IEC 81001-5-1

Security activities within the health-software lifecycle

ISO 13485

Quality management processes supporting consistent conformity

Standards are generally voluntary and should be selected according to the product and its risks. Where a standard is harmonised and cited in the Official Journal, its use may provide a presumption of conformity only for the requirements it covers. The European Commission maintains current information on harmonised medical-device standards.

Further lifecycle considerations are discussed on the MDRC pages covering IEC 62304 compliance and cybersecurity for medical software.

07

Technical documentation should tell one coherent story

The technical file is not a warehouse of disconnected documents. It is a traceable argument showing how the software was conceived, developed, evaluated, released and placed under lifecycle control.

The intended purpose leads to requirements. Requirements lead to risks and controls. Controls lead to verification, clinical evidence and the final benefit-risk conclusion.

Regulatory inputEngineering controlEvidence outputRelease state
Intended purposeSystem requirementsProduct specificationDefined
Foreseeable riskArchitecture and controlRisk-management fileControlled
GSPR requirementVerification methodTest result and trace recordDemonstrated
Clinical claimEvaluation methodClinical or performance evidenceSupported
Released versionConfiguration controlRelease and change recordsIdentifiable
The GSPR matrix is a routing document

It should connect every applicable requirement to specific controlled evidence. A broad statement such as “compliant with IEC 62304” is not a substitute for this traceability.

08

Demonstrate clinical or performance value #

Software can function exactly as designed and still fail clinically. A technically accurate algorithm is not automatically clinically meaningful.

The manufacturer must demonstrate that the relationship between the software output and the targeted clinical condition is valid and that the software performs adequately in its intended population, environment and workflow.

MDR software

Valid clinical association, technical performance and clinical performance support the clinical evaluation.

IVDR software

Scientific validity, analytical performance and clinical performance support the performance evaluation.

The evidence may include literature, analytical studies, retrospective datasets, prospective investigations, usability evidence, real-world data and post-market information. A clinical investigation or performance study is not automatically required for every product. The evidence needed follows from the claims, novelty, risk, inputs, outputs and intended population.

The European Commission’s current guidance collection includes MDCG 2020-1 on clinical and performance evaluation of MDSW. MDRC provides further guidance on clinical evaluation for software-centred medical technologies.

09

Put the regulatory roles, UDI and registration in place #

Every manufacturer must have at its disposal at least one Person Responsible for Regulatory Compliance who meets the Article 15 qualification requirements. Micro and small enterprises may use an external PRRC, but the person must remain permanently and continuously available.

A manufacturer established outside the European Union must designate a sole EU Authorised Representative for at least all products belonging to the same generic device group. The relationship is established through a written mandate accepted by the representative.

PRRC

Regulatory oversight with the qualification and availability required by Article 15

EU representative

Formal EU interface for manufacturers established outside the Union

Medical software is also subject to the UDI system. The manufacturer must establish the applicable Basic UDI-DI and UDI-DI structure, follow the selected issuing entity’s rules and determine how the identifier is presented for the software.

01Actor registrationManufacturer and representative where applicable
02Basic UDI-DIRegulatory grouping identifier
03UDI-DIDevice and software identification
04EUDAMEDUDI and device data submission

A new UDI-DI may be required when a modification changes the software’s original performance, safety, intended use or interpretation of data. The Basic UDI-DI is not the identifier for an individual software version.

The relationship between these identifiers is explained in the MDRC article on UDI-DI, Basic UDI-DI, GTIN and GMN.

Since 28 May 2026, the first four EUDAMED modules have been mandatory. Actor and UDI/device registration therefore belong in the active market-access plan. The current status is set out in the European Commission’s EUDAMED overview.

10

Complete the applicable conformity assessment #

The conformity route follows from the regulation and risk class. Class I software under the MDR and Class A software under the IVDR can generally be self-declared. These routes still require complete technical documentation, a compliant QMS, evidence, risk management, registration and post-market planning.

Self-declaration is not exemption. It means that the manufacturer performs the conformity assessment without routine Notified Body involvement.

01Route selectedSelf-assessment or Notified Body pathway
02Assessment completedQMS and technical documentation as applicable
03Declaration signedEU Declaration of Conformity by the manufacturer
04CE appliedNotified Body number added where required

For higher software classes, a Notified Body assesses the applicable QMS and technical documentation. The Notified Body does not grant the CE marking. It issues the relevant certificate after successful assessment. The manufacturer then draws up the EU Declaration of Conformity and affixes the CE marking. Where a Notified Body is involved, its four-digit identification number accompanies the mark.

Preparation should begin before the formal application. The manufacturer needs a Notified Body whose designation covers the relevant regulation and device or IVD codes. The assessment will examine whether the intended purpose, classification, risks, evidence, software lifecycle and QMS describe one consistent product.

The MDRC page on Notified Body readiness explains how to prepare for that review.

11

Treat CE marking as the beginning of the controlled lifecycle #

Medical software does not stop changing when it reaches the market. Operating systems evolve. Libraries are updated. Vulnerabilities appear. Data distributions shift. Users find unexpected workflows and clinical practice moves forward.

Observe

Complaints, performance, vulnerabilities, trends and real-world use

Evaluate

Clinical impact, risk, regulatory significance and need for action

Control

Correction, update, communication, evidence and configuration

Document

PMS reports, PSUR where applicable, CAPA and change records

Verify

Regression testing, security testing and effectiveness checks

Release

Approved version, updated information and continuing conformity

A patch may be technically small but clinically significant. A model update may improve average performance while introducing a new risk for an underrepresented patient group. A new data source may change the regulatory boundary of the product.

Every change must be evaluated for its effect on intended purpose, classification, safety, performance, cybersecurity, clinical evidence, UDI and certification status.

Conclusion

CE marking is the release of a controlled medical technology

Obtaining CE marking for medical software is not a document-production exercise. It is the process of turning a medical idea into a controlled and evidence-supported product.

The intended purpose defines the promise. Classification reflects the possible consequences when that promise is not fulfilled. Risk management, lifecycle engineering and cybersecurity control how the software is built. Clinical or performance evaluation shows that its output matters. The quality system keeps those elements connected as the software changes.

When these parts are aligned, CE marking becomes more than permission to enter the European market. It becomes evidence that innovation has been translated into a medical technology that can be trusted, monitored and improved throughout its lifecycle.

What manufacturers most often underestimate

The hardest part is rarely one missing document. The real challenge is maintaining alignment across the entire product system.

01

Marketing claims exceed the intended purpose supported by evidence.

02

Classification is selected before the clinical consequences are understood.

03

Requirements cannot be traced to risks, controls and test results.

04

IEC 62304 is treated as proof of complete MDR or IVDR conformity.

05

Algorithm accuracy is demonstrated without showing clinical relevance.

06

Cybersecurity, UDI, EUDAMED and PMS are left until the end.

Software regulatory pathway

Need to align the product build with the regulatory build?

Discuss your software pathway ↗
↑

Discuss your regulatory pathway

Whether you are developing medical software, an AI-enabled healthcare solution, a digital health platform or a software-driven medical device, we can help define the regulatory strategy, evidence requirements and next steps.

→

Where companies need support

Regulatory strategy
Technical documentation
Clinical and performance evidence
Software lifecycle and cybersecurity
Notified Body readiness

Start a discussion

 

Regulatory support for software medical devices, AI health technologies and digital healthcare solutions.


©2026 MDRC - Medical Devices Regulatory Compliance

Useful information

CE-Certificate vs. EC-Certificate

Basic UDI-DI (bUDI)

EUDAMED registration - a brief guide

Authorised Representative Mandate

GSPR – General Safety and Performance Requirements

How to obtain CE marking for medical software under the EU MDR or IVDR?

Technical documentation for Medical Device Software in the EU

Read more >>


Cookie Policy

We only use essential cookies that enable core functionality and proper operation of the website. These cookies do not store any personally identifiable data. By continuing to use this website, you consent to the use of the essential cookies. You may disable these cookies by changing your browser settings, but this may affect how the website functions.
We do not use our own or third-party analytical, preferences, statistics, marketing, functional, advertisement, performance or any other non-essential cookies.

Send us an email:
info@mdrc-services.com

Or use the contact form below

 

Solutions

EU Authorised Representative (EC REP)

EU PRRC

Technical documentation

Risk management

Clinical evaluation

Notified Bodies

Quality management system

Post-market surveillance

Resources

Medical Device Regulation (MDR) - basics

CE-marking process for medical devices

CE-marking process for in vitro diagnostic medical devices

MDR technical documentation checklist

IVDR technical documentation checklist

Technical documentation checklist for medical device software (MDSW)

MDR-compliant quality system documentation checklist

MDR-compliant quality system documentation checklist for medical device software

Clinical Evaluation Plan checklist

Clinical Evaluation Report checklist

PRRC under MDR or IVDR

Articles

CE-Certificate vs. EC-Certificate

Basic UDI-DI (bUDI)

EUDAMED registration - a brief guide

Authorised Representative Mandate

GSPR – General Safety and Performance Requirements

More articles >>

Devices

General medical devices and equipment

In vitro diagnostics (IVD)

Medical software

Cookie Policy

We only use essential cookies that enable core functionality and proper operation of the website. These cookies do not store any personally identifiable data. By continuing to use this website, you consent to the use of the essential cookies. You may disable these cookies by changing your browser settings, but this may affect how the website functions.
We do not use our own or third-party analytical, preferences, statistics, marketing, functional, advertisement, performance or any other non-essential cookies.