MDRC logo
Explore
← Knowledge centre

MDR Article 1(2) · Annex XVI · Regulations 2022/2346 and 2022/2347

Annex XVI devices under the MDR

No medical purpose. No regulatory shortcut.

A coloured contact lens, a dermal filler and a brain-stimulation headset may be sold for very different reasons. The MDR sees the common thread: technology that interacts with the body and can create risks comparable to those of a medical device.

15 min read
Open the regulatory case ↓

The regulatory paradox

The claim may be cosmetic. The harm is still physical.

Annex XVI does not pretend that an aesthetic product has a medical purpose. It does something more precise. It brings six defined product groups under the MDR because their functioning and risks justify medical-device-level controls.

The result is not “MDR-lite”. Qualification, classification, common specifications, technical documentation, clinical evidence, conformity assessment and post-market duties form one connected system.

01

A product does not need to treat disease to create medical-device risk #

Cosmetic contact lenses can injure the cornea. Fillers can cause vascular occlusion or infection. Energy-based equipment can burn, scar or alter pigmentation. Brain stimulation can change neuronal activity. Removing the therapeutic claim does not remove the biological interaction.

Article 1(2) of the MDR therefore extends the Regulation to the product groups listed in Annex XVI from the date the relevant common specifications apply. The legal trigger is not a vague resemblance to healthcare. The product must fit the wording of an Annex XVI group and the scope of the corresponding common specifications.

Medical-purpose side Clinical claim

Diagnosis, prevention, monitoring, treatment or another medical purpose

Comparable technology and riskDifferent purpose · shared control logic
Annex XVI side Non-medical claim

Aesthetic appearance, body modification or another listed non-medical effect

The scope rule

“Aesthetic product” is not itself a legal category. Some aesthetic products fall within Annex XVI, some remain cosmetics or general consumer products, and some are medical devices because their intended purpose is medical.

02

Six doors lead into Annex XVI #

The list is technology- and purpose-specific. It is not an open invitation to regulate every beauty product as a device. Each door has boundaries, inclusions and exclusions.

01EyeNon-corrective contact lenses

Contact lenses or other items intended to be introduced into or onto the eye.

02AnatomyBody-modifying implants

Surgically invasive products for modifying anatomy. Annex XVI also names fixation, while the current common-specification scope excludes it.

03FillingDermal and mucosal fillers

Substances, combinations or items introduced for facial, dermal or mucous-membrane filling, excluding tattooing.

04Adipose tissueFat-reduction equipment

Equipment used to reduce, remove or destroy adipose tissue, including liposuction, lipolysis and lipoplasty systems.

05EnergyHigh-intensity radiation equipment

Lasers, IPL and other high-intensity electromagnetic radiation equipment for specified skin treatments.

06Neural activityBrain-stimulation equipment

Non-invasive equipment using electrical, magnetic or electromagnetic fields that penetrate the cranium.

Details matter. Sunbeds are excluded from the group 5 common specifications. Mesotherapy products fall within group 3 only when their intended purpose includes filling. A laser that removes fat and tightens skin may enter both groups 4 and 5, with the relevant requirements applying cumulatively.

03

Qualification starts with the exact product story #

Do not begin by forcing the product into a risk class. First establish whether it qualifies as an Annex XVI product at all. Read the intended purpose, technology, mechanism, body interaction, user, treatment setting and promotional claims against both the Annex XVI wording and the scope section of Regulation 2022/2346.

01PurposeWhat effect is claimed?

Define the non-medical effect precisely across labelling, instructions, website and sales material.

02TechnologyHow is the effect achieved?

Map energy, material, invasiveness, tissue interaction, duration and accessories.

03Legal wordingDoes it fit a listed group?

Test every element against Annex XVI and the applicable common-specification scope.

Qualification resultIn scope · out of scope · borderline · dual purpose

Dual purpose raises the bar

A device with both medical and non-medical intended purposes must cumulatively fulfil the requirements applicable to both. Where the rules point in different directions, the more stringent requirement may govern. Claims cannot be split into convenient regulatory fragments when the same product and risk controls support both uses.

04

Qualification opens the door. Classification decides the route. #

Once the product qualifies, apply the MDR classification rules in Annex VIII and, for certain active groups, Implementing Regulation 2022/2347. Classification is product-specific, but the following pattern shows why a blanket “aesthetic products are low risk” assumption fails.

Product groupIndicative classRegulatory driver
Contact lenses

IIa or IIb

Duration of use under Rule 5
Body-modifying implants

IIb or III

Implantable, absorbable and specific Rule 8 features
Dermal or mucosal fillers

IIb or III

Implantation, absorption, biological effect or incorporated substance
Fat-reduction equipment

IIb

Reclassification under Regulation 2022/2347
High-intensity skin-treatment equipment

IIa or IIb

Hair-removal-only equipment is IIa, other covered skin treatment is IIb
Brain-stimulation equipment

III

Reclassification under Regulation 2022/2347

Components can follow a different rule from the main equipment. Reusable surgical instruments, cannulas and connecting tubes are examples where the classification analysis must be performed at the level of the actual device, not copied from the headline group.

The conformity consequence

Notified-body involvement follows the applicable MDR class and conformity route. It is common for Annex XVI products, but it should be concluded from classification rather than treated as an automatic premise.

05

Common specifications turn broad MDR principles into product controls #

Implementing Regulation 2022/2346 has one common layer for all six groups and dedicated annexes for each group. It applies from 22 June 2023. The text does more than name hazards. It connects risk management, design controls, information for safety, consumer communication and post-production evidence.

CoreSafe non-medical performance
01Risk management

Lifecycle process, trained personnel, traceability and explicit acceptability criteria

02Design and manufacture

Group-specific controls for biological, mechanical, energy and use-related hazards

03User and consumer

Qualifications, exclusions, treatment context, foreseeable misuse and emergency action

04Information for safety

Expected performance, residual risks, follow-up, warnings and plain-language material

05Post-production loop

Complaints, clinical data, new hazards, trend signals and corrective action

The common specifications must be applied together with the rest of the MDR. They do not replace Annex I GSPRs, the QMS, technical documentation, economic-operator obligations, UDI, PMS, vigilance or the applicable conformity-assessment procedure.

06

No clinical benefit does not mean no clinical evidence #

For Annex XVI products, the MDR requirement to demonstrate clinical benefit is understood as a requirement to demonstrate performance. The clinical evaluation is based on relevant safety data, including PMS, PMCF and, where applicable, a specific clinical investigation.

This changes the centre of gravity of the evidence package. The manufacturer must prove that the promised non-medical effect is achieved and that the associated risk remains acceptable, even though no therapeutic benefit is being claimed.

ClaimExpected non-medical performance

What visible, physical or functional effect does the manufacturer promise?

ExposureReal conditions of use

Who uses it, on whom, how often, in what environment and with which foreseeable errors?

SafetyClinical consequence

What harms occur, how often, how severe are they and how effective are the controls?

LifecycleContinuing evidence

How will PMS and PMCF confirm that the pre-market conclusions remain valid?

Equivalence is a demanding route

Regulation 2022/2346 allows analogous devices with a medical purpose to inform the analysis, but similarity is not a shortcut around access to data, technological comparison, biological comparison and clinical relevance. A neighbouring technology can illuminate risk without proving equivalence.

07

Without medical benefit, risk has nowhere to hide #

A therapeutic device may justify residual risk against expected clinical benefit. A product used only for an aesthetic or other non-medical purpose cannot rely on that same argument. The common specifications require risks to be eliminated or reduced as far as possible and set a deliberately demanding approach to acceptability.

Medical benefitNot claimed
Risk controls carry the case
Required responseReduce risk as far as possible
01Design out the hazard

Start with inherent safety in design and manufacturing.

02Add protective measures

Control what cannot reasonably be eliminated.

03Inform and train

Use warnings and training after stronger controls have been considered.

04Test the whole residual picture

Confirm that one control has not created a different, greater risk.

Risk analysis must include normal and fault conditions, foreseeable misuse, user competence, consumer characteristics and the use environment. For lay use, the distance between a technically correct instruction and actual behaviour becomes a design input.

08

The consumer must see more than the promise #

A glamorous claim can fit in six words. An informed decision needs more. The common specifications require the non-medical purpose, expected performance, residual risks and relevant follow-up to be made visible in a form suited to users and consumers.

Front of regulatory story non-medical purpose

A clear description of the effect the product is intended to achieve

Expected performanceWhat should happen

Describe the non-medical effect in understandable terms.

Residual riskWhat may go wrong

Explain risks, side effects and their control measures clearly.

SuitabilityWho should not use it

State excluded consumer groups, contraindications and special conditions.

AftercareWhat happens next

Give lifetime or resorption information, follow-up and escalation instructions.

Information supplied for a solely non-medical purpose must not carry a clinical-benefit claim. For some product groups, the user must provide a plain-language annex to the consumer before treatment. Training requirements and professional-use restrictions must also be translated into actual distribution and use controls.

09

CE marking is the output of a complete MDR system #

Once in scope, the product enters the MDR machinery. The manufacturer needs more than a risk report and a test certificate. The regulatory system must connect the organisation, product evidence, supply chain and post-market feedback.

Controlled productAnnex XVI device
01QMS

Responsibilities, suppliers, production, change and CAPA

02Technical documentation

Device definition, GSPRs, verification, validation and evidence

03Clinical evaluation

Performance, safety, investigations and PMCF strategy

04Market architecture

Economic operators, registration, UDI, label and declaration

05Post-market control

PMS, vigilance, trend reporting and corrective action

06Conformity assessment

Class-based route and notified-body involvement where required

A defensible dossier flows in one direction
  1. Freeze the intended purpose, claims, users and product boundary
  2. Document Annex XVI qualification and all applicable common-specification scopes
  3. Classify every device and accessory using Annex VIII and Regulation 2022/2347
  4. Map GSPRs and common specifications into design and evidence requirements
  5. Build clinical, usability, biological, technical and safety evidence around real use
  6. Align QMS, technical documentation and post-market plans with the conformity route
  7. Confirm notified-body scope and capacity where third-party assessment is required
  8. Release only when labelling, operators, registration and surveillance are connected

10

Transition dates are conditional routes, not blanket extensions #

The common specifications apply from 22 June 2023. Regulation 2022/2346, as amended by Regulation 2023/1194, creates specific transition routes for certain products already lawfully marketed before that date. The applicable end date depends on the evidence and conformity-assessment route.

Starting conditionLawfully marketed before 22 June 2023

No significant change in design or intended purpose

Clinical investigation routeUp to 31 December 2029

The staged conditions include a started clinical investigation and, from 2028, a written notified-body agreement.

No clinical investigation plannedUp to 31 December 2028

Where notified-body involvement is required, a written agreement is needed from 1 January 2027.

Former MDD certificateClass-dependent MDR dates

A separate route may apply when the detailed Article 2(3) and MDR Article 120 conditions are met.

A new product placed on the market after the common specifications became applicable does not inherit these routes merely because development started earlier. Nor does a transition end date excuse a significant change, a missed milestone or the absence of an eligible pre-2023 market history.

The practical move

Build a dated transition memo for each product. Record the qualifying history, selected route, milestones, significant-change controls, notified-body evidence and the exact provision relied upon.

Official sources

The regulatory outcome

A non-medical purpose changes the evidence question, not the need for control

Annex XVI products sit in an unusual category, but their route is not mysterious. Define the product precisely, qualify it against the legal scope, classify it correctly, then connect common specifications with the full MDR lifecycle.

The most difficult dossiers are rarely defeated by one missing test. They fail when claims, risk, performance, user information and post-market plans describe different products.

Different purpose. Comparable risk. One coherent MDR story.

Annex XVI regulatory strategy

Need to turn an aesthetic product into a defensible MDR pathway?

Discuss your product ↗
↑

Discuss your regulatory pathway

Whether you are developing medical software, an AI-enabled healthcare solution, a digital health platform or a software-driven medical device, we can help define the regulatory strategy, evidence requirements and next steps.

→

Where companies need support

Regulatory strategy
Technical documentation
Clinical and performance evidence
Software lifecycle and cybersecurity
Notified Body readiness

Start a discussion

 

Regulatory support for software medical devices, AI health technologies and digital healthcare solutions.


©2026 MDRC - Medical Devices Regulatory Compliance

Useful information

CE-Certificate vs. EC-Certificate

Basic UDI-DI (bUDI)

EUDAMED registration - a brief guide

Authorised Representative Mandate

GSPR – General Safety and Performance Requirements

How to obtain CE marking for medical software under the EU MDR or IVDR?

Technical documentation for Medical Device Software in the EU

Read more >>


Cookie Policy

We only use essential cookies that enable core functionality and proper operation of the website. These cookies do not store any personally identifiable data. By continuing to use this website, you consent to the use of the essential cookies. You may disable these cookies by changing your browser settings, but this may affect how the website functions.
We do not use our own or third-party analytical, preferences, statistics, marketing, functional, advertisement, performance or any other non-essential cookies.

Send us an email:
info@mdrc-services.com

Or use the contact form below

 

Solutions

EU Authorised Representative (EC REP)

EU PRRC

Technical documentation

Risk management

Clinical evaluation

Notified Bodies

Quality management system

Post-market surveillance

Resources

Medical Device Regulation (MDR) - basics

CE-marking process for medical devices

CE-marking process for in vitro diagnostic medical devices

MDR technical documentation checklist

IVDR technical documentation checklist

Technical documentation checklist for medical device software (MDSW)

MDR-compliant quality system documentation checklist

MDR-compliant quality system documentation checklist for medical device software

Clinical Evaluation Plan checklist

Clinical Evaluation Report checklist

PRRC under MDR or IVDR

Articles

CE-Certificate vs. EC-Certificate

Basic UDI-DI (bUDI)

EUDAMED registration - a brief guide

Authorised Representative Mandate

GSPR – General Safety and Performance Requirements

More articles >>

Devices

General medical devices and equipment

In vitro diagnostics (IVD)

Medical software

Cookie Policy

We only use essential cookies that enable core functionality and proper operation of the website. These cookies do not store any personally identifiable data. By continuing to use this website, you consent to the use of the essential cookies. You may disable these cookies by changing your browser settings, but this may affect how the website functions.
We do not use our own or third-party analytical, preferences, statistics, marketing, functional, advertisement, performance or any other non-essential cookies.